Calibrate voltage
Let the unit sit with output ON for one minute, then connect the DMM voltage input to the power
supply.
Step
Front Panel
SCPI
1
Turn the knob to ‘Cal Voltage’ and press the knob.
CAL:VOLT:LEV MIN
2
Measure the output voltage (low point) with the DMM.
3
Enter the measured value, and press the knob.
CAL:VOLT <
reading
>
CAL:VOLT:LEV MAX
4
Measure the output voltage (high point) with the DMM.
5
Enter the measured value, and press the knob.
CAL:VOLT <
reading
>
6
Read DONE or FAIL on the display.
(wait 10 seconds)
SYST:ERR?
Calibrate current
Let the unit sit with output ON for one minute, then connect a current monitoring resistor across the
output terminals to be calibrated and then connect a DMM across the terminals of the monitoring
resistor.
Step
Front Panel
SCPI
1
Turn the knob to Cal Current and press the knob.
CAL:CURR:LEV MIN
2
Measure the low output current with the DMM (DMM reading / shunt resistance).
3
Enter the measured value, and press the knob.
CAL:CURR <
reading
>
CAL:CURR:LEV MAX
4
Measure the high output current with the DMM (DMM reading / shunt resistance).
5
Enter the measured value, and press the knob.
CAL:CURR <
reading
>
6
Read DONE or FAIL on the display.
Calibrate small current
Allow the unit to sit with output ON for one minute before continuing.
Step
Front Panel
SCPI
1
Turn the knob to Cal Current and press the knob.
CAL:CURR:LEV:LOW MIN
2
Wait approximately 40 seconds for the ‘Cal in Prog’ message to leave the display.