![Keysight Technologies E36100 Series Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/keysight-technologies/e36100-series/e36100-series_operating-and-service-manual_1962203056.webp)
CURRent Subsystem
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <
current
> | MINimum |
MAXimum | UP | DOWN
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [MINimum | MAXimum]
Programs the immediate current level of the instrument's output.See
for details.
This command changes the output immediately. The UP and DOWN parameters increase or decrease
the immediate current by the amount specified by CURRent:STEP. A setting that exceeds the maximum
or minimum rated current will cause error -222 (Data out of range).
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement] <
current
> |
DEFault
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]? [DEFault]
Sets the step size for current programming with the CURRent UP and CURRent DOWN commands. The
query returns a number of the form +#.########E+##.
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <
current
> | MINimum |
MAXimum
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]? [MINimum | MAXimum]
Sets the triggered current level in amps. The query returns a number of the form +#.########E+##.
[SOURce:]CURRent:PROTection:CLEar
Clears an overcurrent protection event.
[SOURce:]CURRent:PROTection:DELay[:TIME] <
time
> | MINimum | MAXimum
[SOURce:]CURRent:PROTection:DELay[:TIME]? [MINimum | MAXimum]
Sets the time (in milliseconds) that the overcurrent protection is temporarily disabled after a current
level change. The query returns a number of the form +#.########E+##.
[SOURce:]CURRent:PROTection:STATe ON|1|OFF|0
[SOURce:]CURRent:PROTection:STATe?
Enables or disables overcurrent protection, which causes the instrument to go into a protected state
when the power supply status is in constant current mode. The query returns 1 (ON) or 0 (OFF).
Содержание E36100 Series
Страница 1: ...Operating and Service Guide E36100 Series DC Power Supplies...
Страница 2: ......
Страница 8: ...Storage Temperature 101 Operating Temperature 101 Environmental Conditions 101...
Страница 18: ......
Страница 39: ......
Страница 52: ......
Страница 71: ......