Triggering Commands
ABORt
Clears any pending delayed trigger and returns the trigger system to idle. If INIT:CONT is enabled, the
trigger system is continuously initiated.
INITiate[:IMMediate]
Cause the trigger system to initiate. When the trigger system is initiated, an event on the specified
trigger source causes the corresponding trigger action on the power supply output. This command
completes one full trigger cycle when the trigger source is IMMediate or initiates the trigger subsystem
when the trigger source is BUS.
INITiate:CONTinuous ON | 1 | OFF | 0
INITiate:CONTinuous?
Enables or disables continuous initiation of the trigger system. If this is disabled, an INIT initiates the
trigger system for only one trigger action. If this is enabled, the trigger system is continuously initiated
and INIT is redundant. The query returns 1 (ON) or 0 (OFF).
*TRG
Generates an event trigger to the trigger system when the trigger system has a BUS (software) trigger as
its trigger source (TRIG:SOUR BUS). If the trigger system is not initiated, the *TRG command is simply
ignored.
TRIGger[:SEQuence]:DELay <
seconds
> MINimum | MAXimum
TRIGger[:SEQuence]:DELay? [MINimum | MAXimum]
Sets the time delay (in seconds, from 0 to 32.767) from the detection of a bus trigger to the start of any
corresponding trigger action on the power supply output. The default is 0, and the query returns a
number of the form +#.########E+##.
TRIGger[:SEQuence]:SOURce BUS | IMMediate
TRIGger[:SEQuence]:SOURce?
Specifies the source from which the instrument will accept a trigger, either a bus (software) trigger or an
internal immediate trigger. At *RST, the bus trigger source is selected. To ensure synchronization when
the bus source is selected, send the *WAI (wait) command. You can also use the *OPC? (operation
complete) query or the *OPC command to indicate when the operation has completed. The query
returns BUS (wait for trigger command) or IMM (continuous triggering.