CALibration:STATe <
state
>,<
code
>
CALibration:STATe?
Unsecures or secures the power supply with the calibration security code Of up to nine digits. The query
returns 1 (ON - unsecured for calibration) or 0 (OFF - secured against calibration).
– The query only returns the state; it does not return the passcode.
– The security code is nonvolatile, and does is not changed by power cycling or *RST.
– The front panel limits the user to entering codes of up to six digits.
– The default passcodes are 36102, 36103, 36104, 36105, and 36106.
– Example: Unlock the instrument to enable calibration: CAL:STATe 1,36102
CALibration:STRing "<
string
>"
CALibration:STRing?
Record or read up to 40 characters of instrument calibration information.
– You can store any information related to calibration, such as date or contact information.
– The power supply must be unsecured before sending a calibration message.
– You can read the string regardless of whether the supply is secured.
CALibration:VOLTage[:DATA] <
value
>
Enters an output voltage value (in volts) that you obtained by reading a DMM.
– You must first select a calibration level (CAL:CURR:LEV) for the value being entered.
– This command can only be used when calibration is unsecured and output is ON.
CALibration:VOLTage:LEVel MINimum|MAXimum
Initiates calibration for a specific voltage level. You must do the MINimum calibration before the
MAXimum, and this command can only be used when calibration is unsecured.