Programming Ranges
The following table shows the maximum voltage, current, and power that can be programmed for each
model. The DEFault voltage is always 0 V.
Model Max Voltage (V) Max Current (A) Max Power (W)
E36102A
6
5.0
30
E36103A
20
2.0
40
E36104A
35
1.0
35
E36105A
60
0.6
36
E36106A
100
0.4
40
Note that the actual programmable values are 3% above these nominal values, but instrument
specifications are only guaranteed up to the values shown in this table.
Table of Programming and Readback Resolutions
Model Voltage Programming Voltage Readback Current Programming Current Readback
Small Current Read-
back
E36102A
.001 V
.001 V
.001 A
.001 A
.001 mA
E36103A
.001 V
.001 V
.001 A
.001 A
.001 mA
E36104A
.001 V
.001 V
.001 A
.001 A
.001 mA
E36105A
.001 V
.01 V
.001 A
.0001 A
.001 mA
E36106A
.001 V
.01 V
.001 A
.0001 A
.001 mA