Use, Maintenance, and Care of the Devices
Cleaning Connectors
3-
c.
Let the alcohol evaporate, then blow the threads dry with a gentle
stream of clean, low-pressure compressed air or nitrogen. Always
completely dry a connector before you reassemble or use it.
3. Clean the Mating Plane Surfaces
a.
Apply a small amount of isopropyl alcohol to a new swab.
b.
Clean the center and outer conductor mating plane surfaces. Refer
. When cleaning a female connector, avoid
snagging the swab on the center conductor contact fingers by using
short strokes.
c.
Let the alcohol evaporate, then blow the threads dry with a gentle
stream of clean, low-pressure compressed air or nitrogen. Always
completely dry a connector before you reassemble or use it.
4. Inspect Each Connector
Inspect the connector again to make sure that no particles or residue are
present.