12
The linearity system specification is calculated using the root sum of the
squares (RSS) method by considering the error caused from the reference
sensor used and the DUT error in the system at 50 MHz. The error specification
caused by the reference sensor and the DUT can be found in their respective
data sheet. The RSS error specification calculation is computed by using the
following equation:
The following example of the DUT is measured at 50Mhz, 20dBm:
– DUT used is 8481A
– Reference sensor used is N8481A
System linearity error specification
DUT error2 Ref sensor error2
+
±
=
System linearity error specification
(3%)2 (0.8%)2
+
±
3.10%
±
=
=
Содержание 8480 Series
Страница 1: ...Keysight 8480 Series Coaxial Power Sensors Operating and Service Manual...
Страница 6: ...6 Keysight 8480 Series Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 12: ...12 Keysight 8480 Series Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 58: ...3 Service 58 Keysight 8480 Series Operating and Service Manual Figure 3 1 Schematic diagram...
Страница 91: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 91 Figure B 2 8481A Option 001...
Страница 92: ...B Bulkhead Assemblies 92 Keysight 8480 Series Operating and Service Manual Figure B 3 8482A and 8482B...
Страница 93: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 93 Figure B 4 8481D...
Страница 95: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 95 Figure B 6 8483A...
Страница 97: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 97 Figure B 8 8487A and 8487D...