![Keysight Technologies 8480 Series Скачать руководство пользователя страница 68](http://html1.mh-extra.com/html/keysight-technologies/8480-series/8480-series_operating-and-service-manual_1961706068.webp)
3
Service
68
Keysight 8480 Series Operating and Service Manual
Figure 3-3
Example of an unacceptable waveform
7
Reassemble the Power Sensor, ensuring that the waveform shown on the
oscilloscope does not change. If the waveform has changed, remove the cover
and readjust the black/white and brown/white wires again.
NOTE
You will find that positioning the wire for switching transients affects the offset.
Go back and forth between the two wires, positioning and repositioning, until
both adjustments are deemed acceptable.
Содержание 8480 Series
Страница 1: ...Keysight 8480 Series Coaxial Power Sensors Operating and Service Manual...
Страница 6: ...6 Keysight 8480 Series Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 12: ...12 Keysight 8480 Series Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 58: ...3 Service 58 Keysight 8480 Series Operating and Service Manual Figure 3 1 Schematic diagram...
Страница 91: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 91 Figure B 2 8481A Option 001...
Страница 92: ...B Bulkhead Assemblies 92 Keysight 8480 Series Operating and Service Manual Figure B 3 8482A and 8482B...
Страница 93: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 93 Figure B 4 8481D...
Страница 95: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 95 Figure B 6 8483A...
Страница 97: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 97 Figure B 8 8487A and 8487D...