3
Service
66
Keysight 8480 Series Operating and Service Manual
FET Balance Procedure
1
Set the power meter as follows:
2
Connect the HIGH GAIN OUTPUT on the modified Power Meter to CHANNEL 1
on the Oscilloscope and set it up as follows:
3
Connect the CHOP OUTPUT on the modified Power Meter to CHANNEL 2 on
the Oscilloscope and set it up as follows:
CAL FACTOR
100%
POWER REF
ON
POSITION
0 Volts (centered)
COUPLING
AC
PROBE
1:1
DISPLAY
AVERAGE
AVERAGE
8
VECTORS
ON
GRID
ON
MAIN/DELAYED
MAIN
TIME REF
CENTER
TIME/DIV
500
μ
s
VOLTS/DIV
50 mV
MODE
AUTO LEVEL
TRIGGER SOURCE
CHANNEL 2
VOLTS/DIV
5 V
POSITION
-5 V
Содержание 8480 Series
Страница 1: ...Keysight 8480 Series Coaxial Power Sensors Operating and Service Manual...
Страница 6: ...6 Keysight 8480 Series Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 12: ...12 Keysight 8480 Series Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 58: ...3 Service 58 Keysight 8480 Series Operating and Service Manual Figure 3 1 Schematic diagram...
Страница 91: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 91 Figure B 2 8481A Option 001...
Страница 92: ...B Bulkhead Assemblies 92 Keysight 8480 Series Operating and Service Manual Figure B 3 8482A and 8482B...
Страница 93: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 93 Figure B 4 8481D...
Страница 95: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 95 Figure B 6 8483A...
Страница 97: ...Bulkhead Assemblies B Keysight 8480 Series Operating and Service Manual 97 Figure B 8 8487A and 8487D...