VFHOMT-11/18 -
11
Maintenance Instructions
Supreme Air Venturi Fume Hoods
CAUTION
!
Periodically:
As Needed:
Prior to performing any maintenance work, verify with the chemical hygiene
officer in charge of the laboratory that there are no hazardous contaminates
on the surfaces inside the hood and which cleaners are compatable with the
chemicals used in the hood.
•
Check the hood for proper face velocity at least annually.
•
Clean glass panel shielding LED lights from hood interior. On hoods with
vapor proof or explosion proof lights, clean the exterior of the globe.
•. Clean both sides of the sash glass.
•. Clean work top and apply a coat of polish.
•
Check sash belts for damage or wear. Replace if frayed. Apply white lithium
grease to the bearings on the sash pulley axles.
•
Plumbing service fittings can be serviced by removing the hood outside
end panels. If the hood is located against a wall or another hood, removable
gasketed access panels are provided (except on Isotope and Perchloric Acid
hoods) in the hood interior. These panels may be removed by pulling gently
on the gasket perimeter until the gasket and panel are free.
•
For hoods with vapor proof lights, the lights can be accessed from the
hood interior. The globe can be unscrewed from the fixture to expose the
incandescent bulb. The fixture is rated for a maximum bulb size of 150 watts.
•
For hoods with explosion-proof light, the light can be accessed from the
hood interior. Remove the light panel by unscrewing the bolts at each corner.
Loosen the set screw and unscrew the globe assembly from the upper
portion of the fixture to gain access to the incandescent bulb. The fixture is
rated for a maximum bulb size of 150 watts.