Step
Operation
Display
Explanation
4
,
or
,
Press
key and input the coating thickness (0.0
µm) of substrate by
,
or
,
keys.
5
Press
key, Zero adjustment is fixed and change
to foil calibration procedure. The word on display is
changing from Standard0 to Standard1.
* If the
Esc
key is selected and
key is pressed during the operation
of steps
2
-
5
then the calibration of the substrate will not be fixed.
6
Measuring the 100 µm
calibration foil
Place calibration foil on
substrate and measure
4-5 times
···
Place the calibration foil (the 100 µm plastic foil) on
the substrate aligning it with the substrate grooves
and measure 4-5 times.
• Depending on the material, a measurement value
that is greatly different from that of the calibration
foil thickness may be displayed, but in steps
7
and
8
this will be fixed as the set value, so proceed
with the operation.
• Measure the calibration foils for calibration in order
of thinnest to thickest.
1
2
5