Step
Operation
Display
Explanation
4
,
or
,
Input the coating thickness(0.0 µm) of substrate by
,
or
,
keys. (Un-coated substrate
means that 0.0 µm coating thickness)
5
Press
key, Zero adjustment is fixed and change
to foil calibration procedure.
* If
Esc
is selected and
key is pressed during the operation of
steps
2
-
5
then the calibration of the substrate will not be fixed.
6
Measuring the 100 µm
calibration foil
Place calibration foil on
substrate and measure
4-5 times
Example in Electromagnetic Mode
···
Place the calibration foil (the 100 µm plastic foil) on
the substrate aligning it with the substrate grooves
and measure 4-5 times.
• Depending on the material, a measurement value
that is greatly different from that of the calibration foil
thickness may be displayed, but in step
7
and
8
this will be fixed as the set value, so just proceed.
• In the target portion of the display, the previously
input calibration foil numeric value is displayed.
1
2
5