(1) Simplified <Example: Calibrate by using a substrate and a calibration foil (100µm foil)>
1
2
5
Step
Operation
Display
Explanation
1
or
Turn the power on by pressing
key, and select
CAL
by
or
keys.
* The display will show the previous measurement mode, either “Fe” or “NFe”.
2
Press
key and the calibration screen will appear.
3
Measuring the substrate
Measure the substrate
4-5 times
Example in Electromagnetic Mode
···
Measure an uncoated substrate 4-5 times. Each
time a measurement is completed a “beep” will
sound and the measurement value will be displayed.
• Electromagnetic measuring mode (Fe) is used
for ferrous substrate samples, and Eddy-Current
measuring mode (NFe) is used for non-ferrous
metal substrate.
• Depending on the substrate, a measurement value
that is greatly different from that shown on the left (a
measurement value close to 0) may be displayed,
but in step
4
and
5
they will be fixed as the set
value, so continue with the operation.