36
• The purpose of measuring the substrate and the calibration foil 4-5 times each is to get the average value.
• When
DEL
is displayed, the latest measured data can be deleted. Select
DEL
by
or
keys, and
press
key.
• When
Esc
is displayed, you can return to step
1
. Select
Esc
by
or
keys, and press
key. If the
calibration procedure is going back before Step
1
, all calibration setting will be nullified.
Step
Operation
Display
Explanation
7
,
or
,
Input the coating thickness of calibration foil (100
µm) by
,
or
,
keys.
8
Press
key and display returns to measuring
mode.
This completes the Simplified calibration operation.
* If
Esc
is selected and
key is pressed during
the operation of step
6
-
8
then the calibration of the
calibration foil will not be fixed.