Cal.Methods (Calibration)
Two types of calibration procedures can be available
(Simplified and Multipoint).
Simplified
x
Calibration by using one point on the
substrate and one calibration foil (see p.34)
Multipoint
x
Calibration by using one point on the substrate
and a maximum of 4 points calibration foils (see p.37)
(1) Follow the steps on p. 15 (1), (2).
(2) Select calibration method by
or
keys, and
press
key. Calibration method is set and display
returns to measuring mode.
(3) If you choose not to change calibration method,
select
Esc
by
or
keys, and press
key.
Display returns to measuring mode without changing
of calibration method.