Kett KB-230 Скачать руководство пользователя страница 5

5

 

Do not damage, heat, or place heavy object on the power cable or connectors.  

Otherwise, it may cause fire or electric shock, etc.

 

When connecting with other devices, first, turn off the power of each unit.

 

Do not connect/disconnect the power plug with wet hands. Otherwise, it may cause an electric 

shock, etc.

 

In the maintenance work, remove the plug from the outlet for safety. Otherwise, it may cause an 

electric shock.

 

To disconnect the plug or connector, make sure to hold the plug or connector, not to pull on the 

cable.

Operating environment for safety use

  This product is designed to be used safely in the following conditions.

  Make sure of the operating environment before use.

 

Indoor use

 

Altitude: up to 2000 m

 

Temperature: 5 to 35°C

 

Relative humidity: environmental temperature up to 31°C, max. 80%; at 35°C, max. 65%

 

Power supply voltage fluctuation: within 10% of nominal voltage

 

Transient overvoltage: within the Overvoltage Category II (conforms to IEC60664-1:2007)

 

Pollution Degree 2 (conforms to IEC60664-1:2007)

Installation and storage

 

Avoid using or storing the unit in a location where it would be exposed to excessively high or low 

temperatures, high levels of humidity, direct sunlight, electromagnetic interference, corrosive gases, 

or large amounts of dust.

 

Place the unit on a flat and stable surface where it will not be subjected to significant vibration dur-

ing use.

 

When moving the unit, never tilt it any more than necessary.

 

Take care never to drop or bump the unit or otherwise allow it to be subjected to strong shocks or 

the application of excessive force.

 

When removing the power cable or other cables, never pull on the cord or cable and instead hold 

the plug or cable connector.

 

Use the supplied power cable and make sure to connect to the protective earth.

 

Do not install the unit in a place where handling the power cable and connectors or turning the pow-

er switch is made difficult.

 

Install the unit at least 10 cm away from the walls so that the ventilation openings provided on the 

rear of the unit are not blocked.

 

When the unit is not to be used over an extended period of time, turn off the power and remove the 

power plug from the outlet.

Cleaning

 

To clean the surface of the unit, wipe off using a soft dry cloth.

 

Do not use abrasive detergent such as paint thinner or benzene when cleaning this product.

 

If measurement sample such as powder, grains, or dust gets into gaps between rotation table, re-

move it to prevent malfunction.

Caution

Burn hazard

Electric shock hazard

Содержание KB-230

Страница 1: ...NIR Moisture Analyzer KB 230 Operating Manual Thank you for purchasing this product Please read the operating manual carefully and use this product properly...

Страница 2: ......

Страница 3: ...e the rotation table 12 6 4 Setting the sampling amount 13 7 Making the calibration curve 14 7 1 Setting the measurement range and the number of samples 14 7 2 Setting the sampling times 14 7 3 Prepar...

Страница 4: ...nd it may cause malfunction fire or electric shock Make sure to ground the power cable With the 2P outlet use a conversion plug to ground In the case of 3P it will be automatically grounded If not gro...

Страница 5: ...igh levels of humidity direct sunlight electromagnetic interference corrosive gases or large amounts of dust Place the unit on a flat and stable surface where it will not be subjected to significant v...

Страница 6: ...o the moisture It uses the measurement principal totally different than drying method or Karl Fischer method 2 2 Parts of main unit Front Rear Power switch Fuse replacing holder Power supply connector...

Страница 7: ...7 2 3 Accessories Sample cell holder For petri dish 90mm Power cable Operating manual First guide Light shielding cover Spare fuse Zero adjustment plate Petri dish 90mm Sample cell...

Страница 8: ...no condensation Power supply 100 120V AC 220 240V AC 50 60Hz Power consumption 50W Dimensions and weight 415 mm W x 370 mm D x 230 mm H 11 kg Maximum sampling amount 2kg Maximum diameter of sample 15...

Страница 9: ...ent TABLE key Used to set ON OFF of the rotation table ZERO key Used for zero adjustment CALC key Used for a making calibration curve EXIT key Used to exit the operation MODE key Used to select the me...

Страница 10: ...adjustment will proceed while displaying the absorbance in real time When the calibration is completed the initial screen will be restored If the power is cycled repeat the operation from step 6 Make...

Страница 11: ...d keep the measuring surface not wrinkled The following plastic bag is recommended Material PE polyethylene PP polypropylene PE PA polyethylene polyamide Thickness 0 05 mm or more 3 Polyethylene sheet...

Страница 12: ...hen the sample moves etc When rotating the table using a 90 mm of Petri dish use the sample cell holder Use caution for falling off of the samples during rotation 6 2 Deciding whether to use the light...

Страница 13: ...nce As the sample is gradually raised the fluctuation of displayed absorbance X1 and X2 will be come stable at a point where the near infrared pass through is stopped Consider the sampling amount at t...

Страница 14: ...balanced in the measurement range 7 2 Setting the sampling times Set the sampling times per moisture Generally three to five times is appropriate However if the shape of sample is not consistent set a...

Страница 15: ...ta as the standard moisture content Set this value in step 5 of 7 6 Optical measurement The official method or a measurement using an infrared moisture analyzer is available Note that the sample canno...

Страница 16: ...Select the channel number The channel selection screen will be displayed Using key move the cursor to the channel number to which the calibration curve is to be registered Then press the ENTER key Pr...

Страница 17: ...7 to enter the following points 0 1 N 0 5 N E X T N 0 6 S A V E E X I T 0 1 N 0 5 S A V E E X I T C A N C E L E X I T Save the calibration data When the standard moisture content of all samples are in...

Страница 18: ...ment channel or during setup the following icon will be displayed in the channel selection screen Icon Measurement Calculating coefficient Status Calibration data is not input or data is insufficient...

Страница 19: ...0 F L O U R Set the sample and press the MEA key The moisture will be displayed after measurement Set the number of display digits and the number of measurement as required Setting the channel pa ram...

Страница 20: ...ent is required after the power is turned on If performing zero adjustment without cycling the power operate the following Attach the zero adjustment plate Open the light shielding cover and set the z...

Страница 21: ...referring to the corresponding pages Display Contents Page 1 C H P A R A M E T E R Setting the channel parameter P 22 2 C H D U P L I C A T E Duplicating a channel P 24 3 C A L I B R A T I O N Making...

Страница 22: ...ter the smoothing of continuous measurement Select the smoothing 0 to 5 using key and press the ENTER key to set D I G I T Select the measurement value display digits Select a display digits from nine...

Страница 23: ...the input content check screen will be displayed After checking the items scrolling with key OK will be selected Press the ENTER key to save the channel parameter The channel name input screen will b...

Страница 24: ...ayed press the ENTER key to overwrite the channel at the duplication destination Press the EXIT key The selection screen for duplication destination channel will be restored Note that when a channel e...

Страница 25: ...e MENU key and enter the pass code Move the cursor to 5 PASS CODE with key Then press the ENTER key to set the code 5 N E W C O D E _ _ _ _ Enter the pass code At the time of shipment 9999 is set Ente...

Страница 26: ...port provided on the rear of the main unit of KB 230 Then tighten the screw Connect the printer cable VZC 54 designated one to the input connector provided on the bottom of the printer and tighten th...

Страница 27: ...mended to replace consumable parts at an interval of two years Do not replace the consumable parts yourself Contact the vendor from which you purchased the unit or the Kett Tokyo sales office or a loc...

Страница 28: ...reflectivity of sample is low and the signal level is be low the requirement When sample is not set on the measurement window Set the sample and make sure that the measure ment value is displayed Whe...

Страница 29: ...absorbance saturated sampling amount While the near infrared from the bottom passes through the sample the displayed value will continue to fluctuate The larger amount of samples will provide more st...

Страница 30: ......

Страница 31: ......

Страница 32: ...arts depicted in this user manual may not appear exactly as their actual counterparts but this does not affect their operation or functionality This user manual was intended to be written as clearly a...