7-7.  Calculating coefficient

From the calibration data input from the optical measurement described in the section 7-6, calculate the 

regression of coefficient.

Step

Display

Operation

3 .

0 1

F L O U R

0 2

×

 Select the channel number and calculate the regression.

Perform (1) to (2) described in P.16.
Using [

]/[

] key, move the cursor to the channel number 

to which the calibration curve is to be registered.
Make sure that the check mark icon 

 is displayed, and 

press the [CALC.] key.

3 . 0 1

r :

0 . 9 9 9 5 5

S e :

0 . 0 3 3 2

0

 Check the result of regression calculation.

The calculation is internally executed. When the result is 
displayed, scroll with [

]/[

] key and check the result.

•   Regression correlation coefficient → r

•   Standard error of regression → Se

•   Coefficient → a0, a1, a2

3 . 0 1

S A V E

N 0 5

C A N C E L

 Register the coefficient.

Move the cursor to [SAVE] with [

]/[

] key, and press the 

[ENTER] key. The coefficient will be registered.

*  The coefficient will not be saved, if the [ENTER] key is 

pressed with [CANCEL] selected with 

[

]/[

]

 key.

3 .

0 1

F L O U R

0 2

×

 Complete registration of coefficient.

When the coefficient is registered, the icon turns off and 
the measurement will be ready.

*  When the [CALC.] key is pressed while selecting the channel 

number with no icon, the calculation result can be displayed 

and checked.

[Channel status icon]

  When changing the measurement channel, or during setup, the following icon will be displayed in the channel selection screen.

 

Icon

Measurement Calculating coefficient

Status

×

×

Calibration data is not input, or data is insufficient.

(For making calibration curve, at least three or more calibration data is 

required.)

×

Effective coefficient is not set.

(The main unit updated the calibration data, but the calculation is not 

done.)

(No icon)

×

Effective coefficient is set.

*  If the effective coefficient is set in any of the following method, the measurement is ready (no icon is displayed).

•   Setting the effective coefficient using PC software

•   Setting the effective coefficient referring to "Setting the channel parameter" (P.22)

•   Making the effective coefficient referring to "7-7 Calculating coefficient" (P.22)

*  When changing the measurement channel, the channel with the icon 

 or 

 cannot be selected.

*  Making calibration curve is available with any icon status (channel can be selected while making a calibration curve). How-

ever, the calibration data or coefficient may have been input even when the channel name is blank. Use caution in selecting 

a channel when making a calibration curve.

7-8.  Calibration curve calculation

  Setting offset value and slope correction value will enable correction of measurement value.

Moisture value (displayed moisture) after correction =

Moisture value before correction x Slope correction value + Offset value

*  For how to set the offset value and slope correction value, refer to the items described in "Setting the channel 

parameter" (P.23).

Содержание KB-230

Страница 1: ...NIR Moisture Analyzer KB 230 Operating Manual Thank you for purchasing this product Please read the operating manual carefully and use this product properly...

Страница 2: ......

Страница 3: ...e the rotation table 12 6 4 Setting the sampling amount 13 7 Making the calibration curve 14 7 1 Setting the measurement range and the number of samples 14 7 2 Setting the sampling times 14 7 3 Prepar...

Страница 4: ...nd it may cause malfunction fire or electric shock Make sure to ground the power cable With the 2P outlet use a conversion plug to ground In the case of 3P it will be automatically grounded If not gro...

Страница 5: ...igh levels of humidity direct sunlight electromagnetic interference corrosive gases or large amounts of dust Place the unit on a flat and stable surface where it will not be subjected to significant v...

Страница 6: ...o the moisture It uses the measurement principal totally different than drying method or Karl Fischer method 2 2 Parts of main unit Front Rear Power switch Fuse replacing holder Power supply connector...

Страница 7: ...7 2 3 Accessories Sample cell holder For petri dish 90mm Power cable Operating manual First guide Light shielding cover Spare fuse Zero adjustment plate Petri dish 90mm Sample cell...

Страница 8: ...no condensation Power supply 100 120V AC 220 240V AC 50 60Hz Power consumption 50W Dimensions and weight 415 mm W x 370 mm D x 230 mm H 11 kg Maximum sampling amount 2kg Maximum diameter of sample 15...

Страница 9: ...ent TABLE key Used to set ON OFF of the rotation table ZERO key Used for zero adjustment CALC key Used for a making calibration curve EXIT key Used to exit the operation MODE key Used to select the me...

Страница 10: ...adjustment will proceed while displaying the absorbance in real time When the calibration is completed the initial screen will be restored If the power is cycled repeat the operation from step 6 Make...

Страница 11: ...d keep the measuring surface not wrinkled The following plastic bag is recommended Material PE polyethylene PP polypropylene PE PA polyethylene polyamide Thickness 0 05 mm or more 3 Polyethylene sheet...

Страница 12: ...hen the sample moves etc When rotating the table using a 90 mm of Petri dish use the sample cell holder Use caution for falling off of the samples during rotation 6 2 Deciding whether to use the light...

Страница 13: ...nce As the sample is gradually raised the fluctuation of displayed absorbance X1 and X2 will be come stable at a point where the near infrared pass through is stopped Consider the sampling amount at t...

Страница 14: ...balanced in the measurement range 7 2 Setting the sampling times Set the sampling times per moisture Generally three to five times is appropriate However if the shape of sample is not consistent set a...

Страница 15: ...ta as the standard moisture content Set this value in step 5 of 7 6 Optical measurement The official method or a measurement using an infrared moisture analyzer is available Note that the sample canno...

Страница 16: ...Select the channel number The channel selection screen will be displayed Using key move the cursor to the channel number to which the calibration curve is to be registered Then press the ENTER key Pr...

Страница 17: ...7 to enter the following points 0 1 N 0 5 N E X T N 0 6 S A V E E X I T 0 1 N 0 5 S A V E E X I T C A N C E L E X I T Save the calibration data When the standard moisture content of all samples are in...

Страница 18: ...ment channel or during setup the following icon will be displayed in the channel selection screen Icon Measurement Calculating coefficient Status Calibration data is not input or data is insufficient...

Страница 19: ...0 F L O U R Set the sample and press the MEA key The moisture will be displayed after measurement Set the number of display digits and the number of measurement as required Setting the channel pa ram...

Страница 20: ...ent is required after the power is turned on If performing zero adjustment without cycling the power operate the following Attach the zero adjustment plate Open the light shielding cover and set the z...

Страница 21: ...referring to the corresponding pages Display Contents Page 1 C H P A R A M E T E R Setting the channel parameter P 22 2 C H D U P L I C A T E Duplicating a channel P 24 3 C A L I B R A T I O N Making...

Страница 22: ...ter the smoothing of continuous measurement Select the smoothing 0 to 5 using key and press the ENTER key to set D I G I T Select the measurement value display digits Select a display digits from nine...

Страница 23: ...the input content check screen will be displayed After checking the items scrolling with key OK will be selected Press the ENTER key to save the channel parameter The channel name input screen will b...

Страница 24: ...ayed press the ENTER key to overwrite the channel at the duplication destination Press the EXIT key The selection screen for duplication destination channel will be restored Note that when a channel e...

Страница 25: ...e MENU key and enter the pass code Move the cursor to 5 PASS CODE with key Then press the ENTER key to set the code 5 N E W C O D E _ _ _ _ Enter the pass code At the time of shipment 9999 is set Ente...

Страница 26: ...port provided on the rear of the main unit of KB 230 Then tighten the screw Connect the printer cable VZC 54 designated one to the input connector provided on the bottom of the printer and tighten th...

Страница 27: ...mended to replace consumable parts at an interval of two years Do not replace the consumable parts yourself Contact the vendor from which you purchased the unit or the Kett Tokyo sales office or a loc...

Страница 28: ...reflectivity of sample is low and the signal level is be low the requirement When sample is not set on the measurement window Set the sample and make sure that the measure ment value is displayed Whe...

Страница 29: ...absorbance saturated sampling amount While the near infrared from the bottom passes through the sample the displayed value will continue to fluctuate The larger amount of samples will provide more st...

Страница 30: ......

Страница 31: ......

Страница 32: ...arts depicted in this user manual may not appear exactly as their actual counterparts but this does not affect their operation or functionality This user manual was intended to be written as clearly a...

Отзывы: