28
13. Error display
The following errors are displayed when there is a problem with the unit or the use condition.
When an error is displayed, check the status and take measures following the procedures
described below.
Error display
Description
Remedy
E-01
System parameter memory
error
Contact the vendor from which you purchased the unit
or the Kett Tokyo sales office or a local Kett office or
sales office for assistance.
E-02
Arithmetic error
E-03
Motor rotation error
E-04
Sensor internal temperature
error
Check the ambient temperature.
E-05
Sample low reflection
(The reflectivity of sample is
low, and the signal level is be-
low the requirement.)
●
When sample is not set on the measurement window
Set the sample, and make sure that the measure-
ment value is displayed.
●
When zero-adjustment plate is set
The lamp may have blown. Contact the vendor from
which you purchased the unit or the Kett Tokyo sales
office or a local Kett office or sales office for assis-
tance.
E-06
Channel parameter not set
Set an appropriate channel parameter.
E-07
Sensor data input error
Contact the vendor from which you purchased the unit
or the Kett Tokyo sales office or a local Kett office or
sales office for assistance.
E-08
Channel parameter memory
error
E-09
Calibration curve data memory
error
E-10
Device data memory error
E-11
Rotation table rotation error