![Kett KB-230 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/kett/kb-230/kb-230_operating-manual_1958550016.webp)
16
1st
sample
7-6. Optical measurement (entering the calibration data)
■
Set the main unit. (
⇒
5. Preparing the main unit (P.10))
■
Perform measurement under the same condition in making calibration curve and regular measure-
ment.
Following the procedures below, enter the channel names and sample calibration data obtained in
section 7-1 to 7-5. The calibration data consists of standard moisture content and absorbance. The
absorbance can be obtained automatically by performing the optical measurement, or it can be direct-
ly input using the numeric keypad.
* Just prior to this operation, it is recommended to perform zero-adjustment. (→ 9. Zero-adjustment (P.20))
Step
Display
Operation
①
P A S S ?
_
_
_
_
●
Pass code input
Press the [MENU] key and enter the pass code using the
numeric keypad.
* At the time of shipment, "9999" is set.
Refer to "Setting the pass code" (P.25).
②
1 . C H
P A R A M E T E R
▶
2 . C H
D U P L I C A T E
3 . C A L I B R A T I O N
▶
4 . C H
R E S E T
●
Select the menu (calibration).
The menu screen will be displayed.
* Pressing the [EXIT] key will restore the initial screen.
Press the [
➡
] key twice. Move the cursor over "3. CALI-
BRATON", and press the [ENTER] key.
③
3 .
0 1
▶
0 2
×
●
Select the channel number.
The channel selection screen will be displayed.
Using [
扌
]/[
➡
] key, move the cursor to the channel number
to which the calibration curve is to be registered. Then,
press the [ENTER] key.
* Pressing the [EXIT] key will restore the menu screen.
* Referring to "Channel status icon" (P.18), avoid overwriting
data unintentionally.
④
0 1
N A M E ?
◀
▶
0 1
N A M E ?
◀
F L O
U
▶
0 1
N A M E ?
◀
F L O U R
▶
●
Enter the channel name.
The channel name input screen will be displayed.
(The cursor is at the left end.)
Input the channel name from the 1st character using the
numeric keypad moving the cursor one by one with the
[Enter] key.
Up to 6 characters can be entered.
After input is done, move the cursor to the right end [
▶
]
using the [ENTER] key. Press the [ENTER] key to set the
channel name.
⑤
0 1 N 0 1
C O N T E N T ?
◀
▶
0 1 N 0 1
C O N T E N T ?
◀
1 3 . 1
▶
●
Enter the standard moisture content.
The standard moisture content input screen will be dis-
played.
Input the 1st standard moisture content using the numeric
keypad. Then, move the cursor to the right end [
▶
] with the
[ENTER] key.
The standard moisture content can be left blank and input
later.
Press the [ENTER] key again, and complete the input of
standard moisture content.
* The standard moisture content can be left blank and input
later. Refer to "7-5 Standard moisture content" (P.15).
Channel 01
What is the standard moisture