Other important information
This section gives information on error messages and instructions for cleaning your scale. It also includes
the declaration of conformity and technical data.
4.1 Error messages
Overload
Reduce the load on the scale or the preload.
Underload
Place weighing pan on the scale and ensure that it can move freely.
Weight reading does not stabilize.
1.Ensure a tranquil environment.
2.Ensure that the weighing pan is free to move.
3.Change the setting of the vibration adapter (Section 3.4).
4.If necessary use the dynamic weighing function (Section 3.4).
Not possible to zero scale
Ensure that zeroing is only carried out in the permissible range and
not under overload or underload conditions.
No calibration/adjustment
Disconnect the power cord plug and then plug it in again. (If operating
on the battery, switch the scale off and then on again.) If the error
message reappears, calibrate/adjust the scale (Section 3.3). If this
does not help, contact your dealer or local representative.
EAROM checksum error
Disconnect the power cord plug and then plug it in again. (If operating
on the battery, switch the scale off and then on again.) If the error
message reappears, contact your dealer or local representative.