![KERN FTB Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/kern/ftb/ftb_operating-instructions-manual_1957256017.webp)
17
3.6 Printing master mode settings
In this block all master mode settings can be recorded with a printer.
Press
key to print out the settings.
3.7 Saving settings and exiting master mode
In the last master mode block you can save your settings and revert to
weighing mode.
Press the
key to exit master mode.
Press the
key to save the settings or the
key to cancel them. The scale then reverts to weighing mode.
3.8 Making a typical setting in master mode
You want to set the readability (resolution) at 0.01kg.
Hold down the
key for about 5 seconds to access master mode,
and confirm that you really want this by briefly pressing
(Yes)
within 3 seconds.
Skip the first master mode block CAL (Calibration, not available
with certified scales) by pressing
(No).
Activate the block for scale settings (Scale) and then the subblock
for resolution (Resolution) by pressing
(Yes) twice.
Press
(No) repeatedly until the desired resolution (0.1kg)
is displayed, then confirm by pressing
(Yes).
Answer the prompt End with
(Yes) as you do not want to
make more settings. If you press
(No), however, you can make
further settings.
Answer the prompt Store? (Save in memory) with
(Yes).
The scale reverts to weighing mode and operates with the new
settings. If you reply with
(No), the changes will not be saved.
LiSt
ENd
StorE ?
MAStEr
CAL
SCALE
rESOLU
=01 Ð
ENd
StorE ?
=00Ð