![KERN FTB Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/kern/ftb/ftb_operating-instructions-manual_1957256010.webp)
10
3
Master mode
In master mode the scale settings can be changed and the various functions activated to adapt the scale
for individual weighing needs.
3.1 Overview and operation
As the acceleration value due to gravity is not the same at every location on earth, each balance
must be coordinated in compliance with the underlying physical weighing principle - to the existing
acceleration due to gravity at its place of location ( only if the balance has not already been adjusted
to the location in the factory). This adjustment process must be carried out during the initial start-
up, after change in location and variation of surrounding temperature. It is also recommendable to
adjust the balance periodically during weighing operation in order to obtain exact measured values.
Calibration
CAL
Settings
SCALE
Interface 1
1FACE1
1FACE2
LiSt
End
ENd
NAStEr
J
J
J
J
J
J
J
J
J
J
J
J
J
J
Section 3.3
Section 3.4
Section 3.5
not docu-
mented
Section 3.6
Section 3.7
Weigh mode
=000Ð
Interface 2
J
*
Hold down
key (about 5 seconds)
Briefly press
key (YES).
*
Confirm the MASTER display
within 3 seconds to enter master mode.
Briefly press
key (NO)
J
J
J
3.2 Adjustment (CAL)