3.3.1 Calibration (adusting) of non verifiable balances
This master mode block is not available with certified scales.
Remove the load from the weighing pan and then press the
key to start the calibration procedure.
The flashing display shows the calibration weight. If desired, the
key can be used to select other calibration weights.
Place the indicated calibration weight on the scale and confirm
with
.
Note:
The calibration procedure can be canceled at any time with
the
key.
Wait until the calibration has been successfully completed
(confirmed by the message done in the display) and the scale
reverts to weighing mode.
3.3 Scale calibration (adjustment)
3.3.2 Calibration (adjusting) of verifiable balances
Switch off the scale.
Remove the scale bottom plate by undoing the Torx TX20 retaining screws .
Important
: In order to remove the bottom plate it is necessary to break the
certification seal
affixed
to it!
Once this seal has been destroyed, the scale must be recertified by an accredited
organization, and a new certification seal must be affixed before the instrument may
be used as a certified scale again!
Hold down the service switch (pushbutton) on the analogprint(indicated by an arrow in the drawings
below, while at the same time switching on the scale. Keep the service switch pressed in until Service
appears in the display.