1-6
BOP HIPWR 031912
FUNCTION GENERATOR CHARACTERISTICS (Continued)
Types of Basic waveforms
Sine (Sinusoidal waveform)
Triangle (Triangular waveform)
Pos. Ramp (Ramp waveform)
Neg. Ramp (Sawtooth waveform)
Square (50% Duty Cycle Pulse)
Level (DC waveform)
Maximum number of points per basic wave-
form
Local: 3933
Count (Number of repetitions)
Range: 1 to 255
For Count = 0 the waveform is exe-
cuted indefinitely until stop com-
mand is sent
• Local: Factory default: when the count is
complete or if execution is stopped using
%
, the output of the unit is set to OFF
(Standby indicator lit). Output must be
enabled (press STANDBY) to execute the
waveform again. (See Table 3-3, WAVE-
FORM STOP for additional settings.)
• Remote: When the count is complete or
the waveform is stopped using the
VOLT:MODE FIX command, the output
remains at last output value.
Waveform Name
10 characters maximum
Waveform Parameters
Type: Voltage or Current
Positive Protection Limit:
(Vd-c or Ad-c)
Negative Protection Limit:
(Vd-c or Ad-c)
Count: number
Initial/Repeating
• Type of waveform, voltage or current, can
be selected only for a new waveform and
consequently applies to all waveform seg-
ments. Changing the type for an existing
waveform is not possible, unless the wave-
form is erased first.
• The Limit values (positive/negative) and
the count number apply to all waveform
segments and can be edited at the begin-
ning of a new waveform or later when mod-
ifying a waveform (see PAR. 3.3.9.7).
• All segments except the last may be set to
run once (Initial) or per the Count setting
(Repeating).
Basic waveform parameters
Sine
: Frequency (Hz), Amplitude (Vp-p or Ap-p), Offset (Vd-c or Ad-c),
Start Phase (°), Stop Phase (°).
Triangular
: Frequency (Hz), Amplitude (Vp-p or Ap-p), Offset (Vd-c or Ad-c), Start
Phase (°), Stop Phase (°).
Ramp (Pos. or Neg.
): Frequency (Hz), Amplitude (Vp-p or Ap-p),
Offset (Vd-c or Ad-c).
Square
: Frequency (Hz), Amplitude (Vp-p or Ap-p), Offset (Vd-c or Ad-c).
Level
: Timing (Sec), Amplitude (Vd-c or Ad-c).
• Local: the Start/Stop phases, available for sine and triangle only, can be set or
edited by modifying the existing segment (see PAR. 3.3.9.7).
• Remote: the start/stop phases are global parameters, applicable to all sine/ trian-
gular waveform segments.
• The Start/Stop points are independent and can have any value between 0° and
360°.
TABLE 1-2. BOP GENERAL SPECIFICATIONS (Continued)
SPECIFICATION
RATING/DESCRIPTION
CONDITION
Содержание BOP-MG
Страница 2: ......
Страница 10: ......
Страница 20: ...x BOP 1K 031912 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 37: ...BOP HIPWR 031912 1 17 FIGURE 1 3 BOP OUTPUT CHARACTERISTICS...
Страница 53: ...BOP HIPWR 031912 2 15 FIGURE 2 4 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 54: ...2 16 BOP HIPWR 031912 FIGURE 2 5 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 64: ......
Страница 132: ......
Страница 192: ......