BOP-1K 031912
4-11
•
%
or
ENTER
- EXIT
FIGURE 4-3. MAIN CALIBRATION SCREEN
3. Press either
!
or
@
to initiate Voltage or Current calibration, respectively. To calibrate
multiple units refer to the Instruction Manual Included with the associated parallel or series
connection cable kit. The LCD shows the explanations for the following function keys:
•
!
- INTERNAL - Selects the voltage calibration of the BOP.
•
@
- GAIN - Starts the calibration of the external gain mode of the power supply.
•
#
- EXTERNAL - Starts the external reference calibration.
•
$
- PROTECT - Starts the protection limits calibration of the power supply.
•
%
or
ENTER
- EXIT - Exit Voltage or Current Calibration
4. Pressing
!
,
@
,
#
, or
$
displays following explanations for the function keys:
CAUTION: IF THE CALIBRATE ZERO
!
FUNCTION IS AVAILABLE, PERFORM THIS
FUNCTION FIRST, AS IT AFFECTS ALL OTHER OUTPUT CALIBRATIONS.
•
!
- Calibrate the Zero adjustment for this operational mode. IF THIS FUNCTION IS
AVAILABLE, PERFORM THIS FUNCTION FIRST, AS IT AFFECTS ALL OTHER OUT-
PUT calibrations.
•
@
- Adjust the MAXIMUM (Positive) for this operational mode.
•
#
- Adjusts the MINIMUM (Negative) for this operational mode
•
$
- PROTECT - Adjusts External Protection limits. (Active for EXTERNAL screen only.)
•
%
or
ENTER
to exit this menu.
VOLTAGE
SOURCE
VOLTAGE*Ref Only* CURRENT
0.0000 0.0000
CALIBRATION MODE
DATE FIELD
Содержание BOP-MG
Страница 2: ......
Страница 10: ......
Страница 20: ...x BOP 1K 031912 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Страница 37: ...BOP HIPWR 031912 1 17 FIGURE 1 3 BOP OUTPUT CHARACTERISTICS...
Страница 53: ...BOP HIPWR 031912 2 15 FIGURE 2 4 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Страница 54: ...2 16 BOP HIPWR 031912 FIGURE 2 5 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Страница 64: ......
Страница 132: ......
Страница 192: ......