ABC 040104
B-13
B.53
[SOURce:]CURRent:PROTection[:LEVel]?
QUERY
CURR:PROT?
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV]? {MIN | MAX}
Long Form: [SOURce:]CURRent:PROTection[:LEVel]? {MIN | MAX}
Return Value: <exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Returns value representing current protection level.
CURR:PROT? returns value set by
CURR:PROT. CURR:PROT?MAX returns maximum current protection value (see Table 1-2). This
value is determined at the factory and cannot be changed by the user. CURR:PROT?MIN returns the
minimum current protection value (always 0). (See example, Figure B-5.)
B.54
[SOURce:]CURRent:PROTection:CLEar
COMMAND
CURR:PROT:CLE
Syntax:
Short Form: [SOUR:]CURR:PROT:CLE Long Form: [SOURce:]CURRent:PROTection:CLEar
Description:
Clears overcurrent condition.
Output programmed to 0V; current limit programmed to small value
(approximately 1% of full scale current rating). Overcurrent Protection Tripped is reset so that
CURR:PROT:TRIP? returns 0. Related Commands: CURR:PROT, CURR:PROT:TRIP?. (See exam-
ple, Figure B-5.)
B.55
[SOURce:]CURRent:PROTection:TRIPped?
QUERY
CURR:PROT:TRIP?
Syntax:
Short Form: [SOUR:]CURR:PROT:TRIP? Long Form: [SOURce:]CURRent:PROTection:TRIPped?
Return Value: 0 or 1.
Description:
Identifies whether overcurrent condition was detected.
0 if overcurrent protection tripped. 1 if
overcurrent protection not tripped. Related Commands: CURR:PROT, CURR:PROT:CLE. (See exam-
ple, Figure B-5.)
B.56
[SOURce:]CURRent:LIMit:HIGH Command
CURR:LIM:HIGH
Syntax:
Short Form: [SOUR:]CURR:LIM:HIGH <exp_value>
Long Form: [SOURce:]CURRent:LIMit:HIGH <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Sets limit for maximum power supply output current.
If the value is out of the acceptable current
range for the power supply model, error message -222,"Data out of range" is posted in the output
queue. If the user tries to set an output current value larger than the CURR:LIM:HIGH setting, a value
corresponding to the current limit will be programmed and error message -301,"Value bigger than
limit" is posted in the output queue. NOTE: If the current limit is changed, any previously established
trigger levels higher than the new limit must be reprogrammed to be lower than the new limit; other-
wise triggers will cause the output to exceed the new limit. Once the limit is established, the unit will
not accept trigger values higher than the programmed limit. Related Commands: CURR. (See exam-
ple, Figure B-5.)
B.57
[SOURce:]CURRent:LIMit:HIGH? Query
CURR:LIM:HIGH?
Syntax:
Short Form: [SOUR:]CURR:LIM:HIGH? Long Form: [SOURce:]CURRent:LIMit:HIGH?
Return Value:<exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Returns value representing current limit set by CURR:LIM:HIGH command.
Related Commands:
CURR:LIM:HIGH, CURR. (See example, Figure B-5.)
B.58
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]
COMMAND
VOLT
Syntax:
Short Form: [SOUR:]VOLT[:LEV][:IMM][:AMP] <exp_value>
Long Form: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude] <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Sets programmed voltage level at power supply output.
This command programs output voltage
to a specific value; actual output voltage will depend on load conditions. If the value exceeds the max-
imum for the model being programmed, error message -222,”Data out of range” is posted in output
queue. If value programmed exceeds the VOLT:LIM:HIGH value, a value corresponding to the voltage
limit will be programmed. Related Commands: *RCL, *SAV, VOLT:LIM:HIGH. (See example, Figure B-
1.
Содержание ABC 10-10DM
Страница 2: ......
Страница 6: ......
Страница 14: ...viii ABC 040104 FIGURE 1 1 ABC SERIES POWER SUPPLY...
Страница 18: ...1 4 ABCOPR040104 FIGURE 1 1 ABC SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING...
Страница 19: ...ABCOPR040104 1 5 FIGURE 1 2 ABC POWER SUPPLY AND RA 71 RACK ADAPTER OUTLINE DRAWING...
Страница 20: ...1 6 ABCOPR040104 FIGURE 1 3 TWO ABC POWER SUPPLIES AND RA 72 RACK ADAPTER OUTLINE DRAWING...
Страница 24: ......
Страница 58: ......
Страница 64: ......
Страница 94: ......
Страница 96: ...FOLD HERE Please place stamp here KEPCO INC 131 38 SANFORD AVE FLUSHING NY 11352 USA CUT HERE...