ABC 040104
B-9
B.32
[SOUR:]LIST:SEQuence:STARt
COMMAND
LIST:SEQ:STAR
Syntax:
Short Form: LIST:SEQ:STAR <int_value> (1 to 40)
Long Form: LIST:SEQuence:STARt <int_value> (1 to 40)
Description:
Determines the first memory location (start of sequence) to be addressed during a pro-
grammed sequence.
Sets memory location (from 1 to 40) of first memory location to be addressed
when a program (chain of memory locations) is executed. This is the start address for the STEP and
RUN commands (LOCal mode). The starting sequence number is stored in non-volatile memory and
is retained when the power supply is turned off. Related Commands: LIST:SEQ:STAR?. (See exam-
ple, Figure B-4.)
B.33
[SOUR:]LIST:SEQuence:STARt?
QUERY
LIST:SEQ:STAR?
Syntax:
Short Form: LIST:SEQ:STAR?
Long Form: LIST:SEQuence:STARt?
Return Value: <int_value) between 1 and 40
Description:
Identifies the starting address for program execution.
Returns the starting sequence location set
by LIST:SEQ:STAR command. Related Commands: LIST:SEQ:STAR. (See example, Figure B-4.)
B.34
[SOUR:]LIST:VOLTage
COMMAND
LIST:VOLT
Syntax:
Short Form: LIST:VOLT[:LEV] <exp_value>
Long Form: LIST:VOLTage[:LEVel] <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Sets the voltage value for the memory location selected by LIST:IND.
If the value is out of the
acceptable voltage range for the power supply model, an error message: -222,"Data out of range" is
posted in the output queue. The voltage value is stored in non-volatile memory and is retained when
the power supply is turned off. Related Commands: LIST:IND. LIST:IND?, LIST:VOLT? (See example,
Figure B-4.)
B.35
[SOUR:]LIST:VOLTage?
QUERY
LIST:VOLT?
Syntax:
Short Form: LIST:VOLT[:LEV]?
Long Form: LIST:VOLTage[:LEVel]?
Return Value: <exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Returns voltage value stored in indexed memory location selected by LIST:IND.
Related Com-
mands: LIST:IND. LIST:IND?, LIST:VOLT. (See example, Figure B-4.)
B.36
[SOUR:]LIST:VOLTage:PROTect
COMMAND
LIST:VOLT:PROT
Syntax:
Short Form: LIST:VOLT:PROT<exp_value>
Long Form: LIST:VOLTage:PROTect <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Sets the voltage protection value for the indexed memory location selected by LIST:IND.
If the
value is out of the acceptable voltage range for the power supply model, an error message:
-222,"Data out of range" is posted in the output queue. The voltage value is stored in non-volatile
memory and is retained when the power supply is turned off. Related Commands: LIST:IND.
LIST:IND?, LIST:VOLT:PROT?. (See example, Figure B-4.)
NOTE: When programming sequential voltage levels, it is important to set the Overvoltage to accommo-
date the highest voltage of the sequence. Otherwise, when going from higher to lower voltage
levels, the overvoltage protection will trip and shut down the unit because the overvoltage setting
registers faster than the power supply can attain the lower voltage.
B.37
[SOUR:]LIST:VOLTage:PROTect
QUERY
LIST:VOLT:PROT?
Syntax:
Short Form: LIST:VOLT[:LEV]?
Long Form: LIST:VOLTage[:LEVel]?
Return Value: <num_value> (digits with decimal point and Exponent)
Description:
Returns voltage protection value stored in indexed memory location selected by LIST:IND.
Related Commands: LIST:IND. LIST:IND?, LIST:VOLT. (See example, Figure B-4.)
Содержание ABC 10-10DM
Страница 2: ......
Страница 6: ......
Страница 14: ...viii ABC 040104 FIGURE 1 1 ABC SERIES POWER SUPPLY...
Страница 18: ...1 4 ABCOPR040104 FIGURE 1 1 ABC SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING...
Страница 19: ...ABCOPR040104 1 5 FIGURE 1 2 ABC POWER SUPPLY AND RA 71 RACK ADAPTER OUTLINE DRAWING...
Страница 20: ...1 6 ABCOPR040104 FIGURE 1 3 TWO ABC POWER SUPPLIES AND RA 72 RACK ADAPTER OUTLINE DRAWING...
Страница 24: ......
Страница 58: ......
Страница 64: ......
Страница 94: ......
Страница 96: ...FOLD HERE Please place stamp here KEPCO INC 131 38 SANFORD AVE FLUSHING NY 11352 USA CUT HERE...