4-2
ABC 040104
NOTES: Keys with dual functions are labeled with both a command and a number. The com-
mand label is referred to when the unit is in command entry status; the number is
referred to when the unit is in data entry status.
1. With the power supply in command entry status , press
CALIB
key. power supply enters
Calibration mode, data entry status .
2. Enter password (4 digits) and press
ENTER
. If password is incorrect,
**WRONG PASS-
WORD**
flashes, then LCD returns to command status. (See PAR. 4.4 to change the pass-
word).
3. If password is correct, LCD reads
VOLTAGE CALIB
. Enter
1
for voltage calibration (see
PAR. 4.3.1), enter
0
to proceed to current calibration.
4. LCD reads
CURRENT CALIB
. Enter
1
for current calibration (see PAR. 4.3.2), enter
0
to exit
Calibration mode.
4.3.1
VOLTAGE CALIBRATION
1. With LCD reading
VOLTAGE CALIB. Connect Instr.
, connect DVM to output (+) and (–
) terminals of power supply. then press ENTER.
2. With LCD reading
Vout ZERO adj.
, monitor DVM and use
"
or
#
keys to increase or
decrease output voltage as necessary to get DVM reading as close to zero as possible (with-
out going to negative value). Press
ENTER
when complete.
NOTE: During calibration the
"
(increase) or
#
(decrease) keys are used for fine adjust-
ments. If necessary, use the
1
(decrease) and
3
(increase) keys for coarse adjustment,
then use the
"
or
#
keys.
3. LCD will flash
Vout F.S.adj. **WAIT**
then
Vout F.S. adj < >
. Monitor DVM and
use
"
or
#
keys to increase or decrease output voltage as necessary to get DVM reading
as close as possible to full scale value (without exceeding full scale value). E.g., for ABC 25-
4DM, adjust for DVM reading as close to 25.000V as possible. Press
ENTER
when com-
plete. If the
"
and
#
keys do not provide the required adjustment range, press the
1
(increase) or
3
(decrease) key for additional range.
4.3.2
CURRENT CALIBRATION
1. LCD reads
CONNECT SHUNT <n OHM >zz A
(n and zz values are determined by ABC
model being calibrated.; e.g., for ABC 10-10DM LCD reads <1 OHM, >50 W). Refer to PAR.
4.2 for shunt requirements. Press
ENTER
.
2. LCD reads
CONNECT DVM TO SHUNT
. Connect Precision Shunt across output (+) and (–)
terminals of power supply and connect DVM to sense terminals of shunt. Press
ENTER
when connections are complete.
NOTE: During calibration the
"
(increase) or
#
(decrease) keys are used for fine adjust-
ments. If necessary, use the
1
(decrease) and
3
(increase) keys for coarse adjustment,
then use the
"
or
#
keys.
3. With LCD reading
Iout ZERO adj.
, monitor DVM and use
"
or
#
keys to increase or
decrease output current as necessary to get DVM reading as close to zero as possible (with-
out going to negative value). Press
ENTER
when complete.
Содержание ABC 10-10DM
Страница 2: ......
Страница 6: ......
Страница 14: ...viii ABC 040104 FIGURE 1 1 ABC SERIES POWER SUPPLY...
Страница 18: ...1 4 ABCOPR040104 FIGURE 1 1 ABC SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING...
Страница 19: ...ABCOPR040104 1 5 FIGURE 1 2 ABC POWER SUPPLY AND RA 71 RACK ADAPTER OUTLINE DRAWING...
Страница 20: ...1 6 ABCOPR040104 FIGURE 1 3 TWO ABC POWER SUPPLIES AND RA 72 RACK ADAPTER OUTLINE DRAWING...
Страница 24: ......
Страница 58: ......
Страница 64: ......
Страница 94: ......
Страница 96: ...FOLD HERE Please place stamp here KEPCO INC 131 38 SANFORD AVE FLUSHING NY 11352 USA CUT HERE...