MouseSTAT Troubleshooting
Problem
Possible Cause
Solutions
There is no red light
or
blinking light in the
sensor.
Sensor not fully connected
• Ensure that all connections are tight,
especially at the back of the controller.
Damaged wire in sensor or
interface cable
• Replace the damaged cable.
Readings are
inconsistent.
or
There are no
readings.
Incorrect sensor placement
• Reposition sensor on the animal’s paw.
Poor animal circulation
• Investigate types and dosages of anesthetic
and other drugs in use; reduce where
possible.
• Increase animal body temperature to
encourage circulation.
MouseSTAT Sensor Maintenance
• Avoid any practice that creates tension or strain on the sensor cable or
connections. MouseSTAT sensors are designed to be gentle on small paws.
• When the MouseSTAT sensor is not in use, gently coil it and store it in its package
to prevent tangles or bends in the sensor wire.
• Gently wipe the sensor clean between uses.