Set Up Hardware
Please see the SomnoSuite instructions on page 3.
Scavenging:
Connect a charcoal canister to the Exhaust port on the back of the SomnoSuite.
Air Source:
If you are using a Compressed Gas Source:
1. Connect the Compressed Gas tubing directly to the Compressed Gas port on the
back of the SomnoSuite.
2. Set gas pressure to 10-15 PSI.
If you are using the Internal Pump:
No additional hardware setup is necessary.
Tubing Configurations:
Connecting an Induction Chamber and Ventilator Tubing to an Endotracheal Tube:
1. Connect one 2- or 3-Accessory Connector to the INSP port.
2. Connect the other 2- or 3-Accessory Connector to the EXP port.
3. Designate a clip color to each accessory, each with an INSP and EXP branch.
4. Connect one INSP-EXP set to the intubation connector tubing, mask tubing,
and induction chamber.
C
F
C
F
Tip: If your gas source does not regulate to this range, connect
a Kent Scientific Pressure Reducer between the gas source and
the SomnoSuite. See Setup on page 4.