9
2) Touch Setup > Air Source to choose the carrier gas source.
Room Air: Utilizes the SomnoSuite’s internal pump to draw in surrounding air
Compressed gas: Utilizes a 10-15 PSI gas source connected to the Compress
Gas Port.
3) Install the filled syringe into the syringe block holder, tightening the Anesthetic
Delivery tube to the syringe tip.
4) Be sure that the L—bracket is tightened over the metal collar of the syringe
rather than the glass barrel.
Tip:
Use
Remove
to move the block up if necessary.
1) Touch Back, then Return to view the Anest Run Screen.
2) Touch and hold Prime to move the block towards the syringe plunger.
Stop when the block touches the top of the plunger.
CAUTION
Do not overprime the delivery tube! Do not allow the pusher
block to depress the syringe plunger or inject liquid anesthetic
into the vaporizer block during this step.
Turn Dial B while holding the Prime button to adjust priming
speed as needed.