The RoVent Run Screen:
RoVent Run Screen
Initial Setup
Ventilator Calibrations:
The first time the RoVent is used, or any time the accessories have changed, be sure to
perform a Deadspace Calibration and verify the ventilator settings.
Note: Never perform these functions when an animal is connected to the ventilator.
1.
Measure Deadspace
The Deadspace calibration will measure the volume of tubing between the
SomnoSuite RoVent module and the animal. It is important that the
Deadspace is recalibrated every time the tubing configuration is changed
to maintain accurate tidal volume delivery.
1) Set up tubing.
2) Connect the cap to the anesthetic delivery tube.
3) From the Vent Run Screen, touch Setup then
More Ventilator
.
4)
Touch
Calib & Tests
, then
Measure Deadspace
.
5) Firmly seal the end of the endotracheal tube connector with your finger
or a cap, ensuring that there is no air leakage.
6) Press Dial B. When asked to confirm measurement, press Dial B again.