Technical data
49
Part / function
Description
Start "start/stop"-key
Stop "start/stop"-key
Short-term start and stop, respectively
„Quick run“-key : short-term run when pressed permanently; stop when
released
Indication of operating state
Preset values for speed and run time ar displayed during the run.
End of centrifugation
The panels "speed" and "time" read "0".
parameter memory,
parameter display
•
speed
•
run time
Speed selection
adjustable in steps of 100 min
-1
in the range of 1600 min
-1
to 13000 min
-1
Run time selection
adjustable in minutes from 1 min to 99 min;
„hd“-mode: continuous operation
Time display in „quick run“ mode
between 1 s and 60 s in seconds, over 1 min in minutes
diagnostics
•
lid not properly closed: display "OPEN";
•
indication of general malfunctions (ERROR codes)