The control panel of the
Biofuge haemo
Display
Speed
Resting state: upon pressing any key: display of
preselected speed, after 2 min switch
to standby mode (flashing red light)
During run:
preselected speed
Braking: "br"
End: "0"
Lid open:
"OPEN"
Resting/end:
error codes (if present)
Run time ("time")
Resting/end:
upon pressing any key: display of
preselected run time, after 2
min
switch to standby mode (flashing red
light)
During run:
remaining run time or (with quick start)
run time passed
Keys
lid:
open lid (only possible if connected
with mains supply)
quick run:
short-term acceleration as long as key
is pressed, with indication of run time
passed
start/stop:
normal start or manual stop
set
speed/time: selection of the parameter to be
changed (speed or run time)
+/-
stepwise increase/decrease of preset
values for the selected parameter
speed or run time, accelerated
change when pressed permanently
Error codes
(troubleshooting see chapter "Troubleshooting")
E-1/E-12: keys do not respond (main board defective)
E-3:
no braking current: lid popped open or
manually opened during run; excess tem-
perature in the drive
br:
power turned off during run or power failure
lid:
lid opened during run
OPEN:
lid lock mechanism out of order
Содержание Heraeus Biofuge haemo
Страница 1: ...Instructions for use Biofuge haemo ...
Страница 3: ...Biofuge haemo set set 1 2 0 0 0 1 0 speed time lid quick run start stop INSTRUMENTS ...
Страница 10: ...For your safety 6 for your notes ...
Страница 18: ...Accessories 14 for your notes ...
Страница 20: ...Before use 16 for your notes ...
Страница 41: ...Operation 37 for your notes ...
Страница 51: ...Troubleshooting 47 ...
Страница 57: ...53 Autoclaving protocol only fixed angle rotor 7500 3325 Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...