![Kendro Heraeus Biofuge haemo Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/kendro/heraeus-biofuge-haemo/heraeus-biofuge-haemo_instructions-for-use-manual_663806034.webp)
Operation
30
Selecting the run time
You can select a run time between 1 and 99 min or
continuous operation.
For hematocrit determinations you
should select a run time of 10 min for a
standard blood column of about
65 mm.
The evaluation of hematocrit measurements is detailed
on page 35.
Preselected run time
To predetermine the run time, proceed as follows:
1. Press the "set" key in the "time" panel (press twice if
the centrifuge is in the standby mode). The run time
last entered appears in the display. The right-hand
digit (in the example on the following page the figure
"5") flashes, thus indicating that you can change the
setting in minutes.
2. Press one of the "set" keys "+" (increase) or "-" (de-
crease). By pressing the selected key
briefly
, you
increase or decrease the preset run time in steps of
1 min. This option is supposed to be used for small
changes and fine tuning.
3. If you keep the key pressed, the display changes at
first slowly and after a few seconds at an acceler-
ated pace.
4. Release the key as soon as you are close to the
desired value and fine tune if necessary by repeat-
edly pressing the selected key (or its counterpart if
you have proceeded too far). The display keeps
flashing. The run time is stored by again pressing
one of the "set" keys or "start".
Continuous operation
For continuous operation, press the "set" key in the
"time" panel and then the "-" key until "hd" (for "hold")
appears in the display.
With this setting, the centrifuge keeps running until
stopped manually.
Please note that the lifetime of centrifuge tubes, par-
ticularly plastic ones, is limited, and that they may be
damaged by continuous operation!
set
set
0
1 5
speed
time
Содержание Heraeus Biofuge haemo
Страница 1: ...Instructions for use Biofuge haemo ...
Страница 3: ...Biofuge haemo set set 1 2 0 0 0 1 0 speed time lid quick run start stop INSTRUMENTS ...
Страница 10: ...For your safety 6 for your notes ...
Страница 18: ...Accessories 14 for your notes ...
Страница 20: ...Before use 16 for your notes ...
Страница 41: ...Operation 37 for your notes ...
Страница 51: ...Troubleshooting 47 ...
Страница 57: ...53 Autoclaving protocol only fixed angle rotor 7500 3325 Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...