Operation
28
Placing the tubes in the rotor
The rotor must be loaded symmetrically. When loading
the rotor only partially, you must ensure that opposite
bores always receive tubes of equal weight (when cen-
trifuging a single sample, place a centrifuge tube e.g.
filled with water). The following figure gives examples
for proper loading.
Improper loading can in the worst case
lead to damage to rotor and centrifuge.
Unbalance not only causes a noisy
run, but rapidly damages the motor
suspension.
After loading the rotor, snap on the rotor cap until it
locks into place. The optional hermetic lid 75003326 is
screwed onto the rotor with the central nut. Close the
lid of the centrifuge by firmly pressing it down. There
must be a clicking sound, and the lid must be locked so
that it cannot be opened manually.
improperly loaded rotors
properly loaded rotors