Troubleshooting
45
Error
Behavior of the centrifuge
Possible cause(s) and measures to be taken
Centrifuge is exceptionally
noisy.
1. Stop the centrifuge by pressing the "stop" key, in case of
emergency pull mains plug.
2. Wait until the centrifuge stands still.
3. Check whether the rotor is properly loaded.
4. Check whether a broken vessel, damage to the rotor or
motor malfunction was responsible for the noise.
5. If you cannot locate and solve the problem, call Service.
br
Mains connection interrupted
during run
If the plug was pulled inadvertently, plug in again.
Wait for about 75 seconds. The centrifuge comes to a stop
without braking.
E-1 or E-12
Rotor does not start
Pull mains plug and plug in again.
If the error persists, call Service.
E-3
No braking current; drive out
of order, rotor stops without
braking
Lid opened during run.
Wait until the rotor has come to a complete stop (check
window in the lid). Pull mains plug and plug in again. If error
persists, call nearest Service.
Motor overheated.
Room temperature too high or insufficient air cooling. Pull
mains plug and wait 30 min. Check all vents and clean with
small brush.
Содержание Heraeus Biofuge haemo
Страница 1: ...Instructions for use Biofuge haemo ...
Страница 3: ...Biofuge haemo set set 1 2 0 0 0 1 0 speed time lid quick run start stop INSTRUMENTS ...
Страница 10: ...For your safety 6 for your notes ...
Страница 18: ...Accessories 14 for your notes ...
Страница 20: ...Before use 16 for your notes ...
Страница 41: ...Operation 37 for your notes ...
Страница 51: ...Troubleshooting 47 ...
Страница 57: ...53 Autoclaving protocol only fixed angle rotor 7500 3325 Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10 ...