Operating manual PT 11x, 12x,13x
22
tomatically and producing temperature readings cyclically) as long as
Limit 2 is exceeded for the time period defined by
T.DEL.
Parameter
Function
Auto reset
(
A.RST
):
Auto reset on/off
Parameter Set Li2 check on tAct [
CH.L.2
]
When this parameter is set to ON (
CH.L.2
=on), the detected tempera-
ture must exceed Limit 2 during the entire measuring time (
TACT
) in
order for the pyrometer to generate a valid temperature reading. The
measurement is discarded when the value falls below Limit 2.
Parameter
Function
Li2 check on tAct
(
CH.L.2
)
on/off
14
Setting Parameters at the device
In addition to the configuration possibilities described, many parameters
can be adjusted at the rear panel using push buttons. These settings can
be accessed via configuration layers.
14.1 Configuration layers
The configuration layers are structured as follows:
C001
Temperature measurement lamdba 1
C010
I/O configuration (LED / Buzzer)
C011
General function
C020
Display temperature readings
The following chart lists all parameters. In the factory default configura-
tion, certain parameters and configuration layers will not be accessible
by control key. This is meant to simplify pyrometer operation for the user.
These parameters are indicated by
. If required, remote access to these
parameters can be enabled via PC interface or at configuration layer
C011
. Open the user calibration menu (Menu mode: Full).
Some parameter settings will be hidden if their basic function has been
deactivated. For example, you will not be able to adjust the smoothing
time of the filter if the filter has been turned off or switched to automatic
mode.
Содержание ITS CellaPort PT 120
Страница 1: ...Operating manual PT 11x 12x 13x Pyrometer CellaPort PT 11x 12x 13x Item No 106 3102 01 2019...
Страница 6: ...Operating manual PT 11x 12x 13x...
Страница 64: ...Operating manual PT 11x 12x 13x 58 33 Dimensions...
Страница 69: ...Operating manual PT 11x 12x 13x 63...
Страница 70: ......
Страница 71: ......
Страница 72: ......