The DCX logger can be connected locally to a laptop via the interface converter
cable.
The Logger 5 software can now be used to configure the device or to transmit the
measurement data stored in the DCX to the laptop.
→
Seal the read-out plug with the protective cap
Water level configuration
The surface of the interface plug integrated into the DCX represents the sys
-
tem
’
s reference level. The measuring system is calibrated during installa
-
tion. For this purpose, the run-off distance is measured (reference level-
surface of water, e.g. using a light plummet) and is entered in the water level
configuration in the Logger 5 software together with the density of the medium.
The system length is the sum of the hydrostatic water pressure measured at the
level sensor and the distance from the reference level that is entered.
After completing work on the measuring point, close the level tube seal by screw
-
ing in the hexagon socket.
→
Allen wrench, size 5 (not included in the scope of delivery)
Tip:
To give the measuring point better protection against access, the Allen screw head
can be replaced by a screw head that is not available through usual commercial
outlets.
System lengt
h
Air Ga
p
Level
Water
surface
Sensor
reference line
Reference level