17
keller-druck.com
To replace the battery, first unscrew the knurled nut and then pull off the battery sleeve.
Carefully disconnect the battery from the plug and replace it.
After changing the battery, guide the plug into the opening in the sleeve, push the bat-
tery in and fix it in the clip. Check the sealing rings. Re-fit the battery sleeve, knurled
nut and fixture.
Check the configuration (time, measurement interval, etc.) of the DCX via the Logger 5
software before using the product.
Watertightness
Products in the Series DCX logger are dustproof and are protected against the effects
of immersion in water (according to the pressure range).
The products are also protected against water splashes in the area of the read-out com
-
ponent.
Exceptions: AA products: may be flooded for brief periods
DCX-22:
continuous/long-lasting immersion
→
Please consult the datasheet for the precise specifications.
Check that all sealing rings are clean and undamaged, and that the battery compart
-
ment cover is firmly closed so that the Series DCX device is watertight.
Rechargeable battery
Type DCX-18 is powered by a rechargeable battery which can only be charged using
interface cable K-114M. The rechargeable battery is welded into the device and cannot
be exchanged. The status of the rechargeable battery is determined by measuring its
voltage. Avoid letting it discharge to a value of less than 10%. If the DCX is not being
used, recording should be stopped and the rechargeable battery should be recharged
once per year to prevent total discharge.
Type of
charging
Supply
System length
Charging current
(rechargeable
battery)
Charge time
(full charging)
normal
USB
max. 20 m
90 mA
~ 8.5 h
fast
Power unit
max. 100 m
790 mA
~ 1 h
Vor öffnen Datenlogger trocknen! Keep dry before opening!
Battery plug
Sealing rings
Battery sleeve
Knurled nut