14
keller-druck.com
Configuring and reading the data logger
The DCX is configured and read locally. The DCX is fitted in the measuring point and is
connected to a laptop via the interface converter cable.
• The Logger 5 Software Manual is available at www.keller-druck.com
Installing the software
Only connect the DCX and the converter cable to your computer after the software has
been installed
• Install the KELLER “Driver K-104 K-114� software
• Install the KELLER “InstallerLogger5� software
→
The connection must not be broken during communication with the DCX (configura-
tion, data read-out). Always close the software program first, and only then unplug
the DCX connection.
DCX time
The internal time for the DCX
is set by synchronization with
your computer’s (PC) time. To do this, enable the “Synchronize device time with PC
time“ field in the Logger 5 software and then register the configuration on the DCX. The
DCX’s internal clock does not take account of summer time/winter time changeovers.
UTC (Universal Time Coordinated)
It is advisable to set the computer’s time zone to UTC.
Since UTC does not implement time corrections, the DCX’s measuring times remain
unchanged throughout the year and they can be converted to local time if necessary.
Time zones
If your computer’s time is assigned to a time zone (e.g. UTC +2), the PC clock will
change over automatically between summer and winter time.
Time corrections / changeovers of the DCX’s internal clock only take place if it is
synchronized with the computer (“Write configuration�).
Tip:
Configurations are only sent to the logger when you press the “Write configuration�
button. You can read the device configuration into the Logger 5 software again to check
the DCX settings.