A-3
METER SPECIFICATIONS
VOLTAGE MEASUREMENT ACCURACY
1
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C):
±(0.15
×
accuracy specification)/°C.
CURRENT MEASUREMENT ACCURACY
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C):
±(0.15
×
accuracy specification)/°C.
CONTACT CHECK
4
ADDITIONAL METER SPECIFICATIONS
LOAD IMPEDANCE:
Stable into 10,000pF typical.
COMMON MODE VOLTAGE:
250VDC.
COMMON MODE ISOLATION:
>1G
Ω
, <4500pF.
OVERRANGE:
101% of source range, 102% of measure range.
MAXIMUM SENSE LEAD RESISTANCE:
1k
Ω
for rated accuracy.
SENSE INPUT IMPEDANCE:
>10G
Ω
.
NOTES
1. Add 50µV to source accuracy specifications per volt of HI lead drop.
2. Four-wire remote sense only.
3. Applies when in single channel display mode.
4. Includes measurement of SENSE HI to HI and SENSE LO to LO contact resistances.
SPEED
MAXIMUM MEASUREMENT
TIME TO MEMORY
FOR 60Hz (50Hz)
4
ACCURACY (1 Year)
23°C ±5°C
±(%rdg. + ohms)
FAST
1 (1.2) ms
5% + 10
MEDIUM
4 (5) ms
5% +
1
SLOW
36 (42) ms
5% + 0.3
RANGE
DISPLAY
RESOLUTION
3
VOLTAGE
BURDEN
2
ACCURACY (1 Year)
23°C ±5°C
±(% rdg. + amps)
100.000 nA
1 pA
<1 mV
0.05 % + 100 pA
1.00000 µA
10 pA
<1 mV
0.025% + 300 pA
10.0000 µA
100 pA
<1 mV
0.025% + 1.5 nA
100.000 µA
1 nA
<1 mV
0.02 % + 25 nA
1.00000 mA
10 nA
<1 mV
0.02 % + 200 nA
10.0000 mA
100 nA
<1 mV
0.02 % + 2.5 µA
100.000 mA
1 µA
<1 mV
0.02 % + 20 µA
1.00000 A
10 µA
<1 mV
0.03 % + 1.5 mA
3.00000 A
10 µA
<1 mV
0.05 % + 3.5 mA
RANGE
DISPLAY
RESOLUTION
3
INPUT
RESISTANCE
ACCURACY (1 Year)
23°C ±5°C
±(% rdg. + volts)
100.000 mV
1 µV
>10 G
Ω
0.015% + 150 µV
1.00000
V
10 µV
>10 G
Ω
0.015% + 200 µV
6.00000
V
10 µV
>10 G
Ω
0.015% +
1 mV
40.0000
V
100 µV
>10 G
Ω
0.015% +
8 mV
2601
System SourceMeter
®
2602
Multi-Channel I-V Test Solutions
GENERAL
HOST INTERFACES:
Computer control interfaces.
IEEE-488:
IEEE-488.1 compliant. Supports IEEE-488.2 common commands and status model
topology.
RS-232:
Baud rates from 300 bps to 115200 bps. Programmable number of data bits, parity
type, and flow control (RTS/CTS hardware or none). When not programmed as the active
host interface, the SourceMeter can use the RS-232 interface to control other
instrumentation.
EXPANSION INTERFACE:
The TSP-Link expansion interface allows TSP enabled instruments to
trigger and communicate with each other.
Cable Type:
Category 5e or higher LAN crossover cable.
Length:
3 meters maximum between each TSP enabled instrument.
DIGITAL I/O INTERFACE:
Connector:
25-pin female D.
Input/Output Pins:
14 open drain I/O bits.
Absolute Maximum Input Voltage:
5.25V.
Absolute Minimum Input Voltage:
–0.25V.
Maximum Logic Low Input Voltage:
0.7V, +850µA max.
Minimum Logic High Input Voltage:
2.1V, +570µA.
Maximum Source Current (flowing out of Digital I/O bit):
+960µA.
Maximum Sink Current @ Maximum Logic Low Voltage (0.7V):
–5.0mA.
Absolute Maximum Sink Current (flowing into Digital I/O pin):
–11mA.
5V Power Supply Pin:
Limited to 600mA, solid state fuse protected.
Output Enable Pin:
Active high input pulled down internally to ground with 10k
Ω
resistor.
When the Output Enable input function has been activated, each SourceMeter channel will
not turn on unless the Output Enable pin is driven to >2.1V (nominal current = 2.1V /
10k
Ω
= 210µA).
POWER SUPPLY:
100V to 240VAC, 50–60Hz (manual setting), 240VA max.
COOLING:
Forced air. Side intake and rear exhaust. One side must be unobstructed when rack
mounted.
WARRANTY:
1 year.
EMC:
Conforms to European Union Directive 89/336/EEC, EN 61326-1.
SAFETY:
Conforms to European Union Directive 73/23/EEC, EN 61010-1, and UL 61010-1.
DIMENSIONS:
89mm high
×
213mm wide
×
460mm deep (3
1
⁄
2
in
×
8
3
⁄
8
in
×
17
1
⁄
2
in). Bench
Configuration (with handle & feet): 104mm high
×
238mm wide
×
460mm deep (4
1
⁄
8
in
×
9
3
⁄
8
in
×
17
1
⁄
2
in).
WEIGHT: 2601:
4.75kg (10.4 lbs).
2602:
5.50kg (12.0 lbs).
ENVIRONMENT:
For indoor use only.
Altitude:
Maximum 2000 meters above sea level.
Operating:
0°–50°C, 70% R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C.
Storage:
–25°C to 65°C.
+5VDC
5.1k
Ω
100
Ω
600mA
Solid State
Fuse
Read by
firmware
Written by
firmware
+5V Pin
(on DIGITAL I/O
connector)
Digital I/O Pin
(on DIGITAL I/O
connector)
GND Pin
(on DIGITAL I/O
connector)
Rear Panel