A-7
Series 2600
System SourceMeter
®
Multi-Channel I-V Test Solutions
SPEED SPECIFICATIONS
1
MAXIMUM SWEEP OPERATION RATES (operations per second) FOR 60Hz (50Hz):
MAXIMUM SINGLE MEASUREMENT RATES (operations per second) FOR 60Hz (50Hz):
A/D CONVERTER
SPEED
TRIGGER ORIGIN
MEASURE
TO GPIB
SOURCE MEASURE
TO GPIB
SOURCE MEASURE
PASS/FAIL
TO GPIB
0.001 NPLC
Internal
1110 (1000)
880 (880)
840 (840)
0.01 NPLC
Internal
950 (900)
780 (760)
730 (710)
0.1
NPLC
Internal
390 (345)
355 (320)
340 (305)
1.0
NPLC
Internal
57 (48)
56 (47)
56 (47)
A/D CONVERTER
SPEED
TRIGGER ORIGIN
MEASURE
TO MEMORY
MEASURE
TO GPIB
SOURCE MEASURE
TO MEMORY
SOURCE MEASURE
TO GPIB
SOURCE MEASURE
PASS/FAIL
TO MEMORY
SOURCE MEASURE
PASS/FAIL
TO GPIB
0.001 NPLC
Internal
10000 (10000)
8000 (8000)
5500 (5500)
3600 (3600)
4900 (4900)
3100 (3100)
0.001 NPLC
Digital I/O
2700 (2650)
2100 (2100)
2300 (2300)
1900 (1875)
2200 (2150)
1800 (1775)
0.01 NPLC
Internal
4000 (3500)
3600 (3200)
2750 (2700)
2300 (2100)
2800 (2500)
2100 (1975)
0.01 NPLC
Digital I/O
1900 (1775)
1600 (1500)
1700 (1600)
1450 (1400)
1600 (1500)
1400 (1325)
0.1
NPLC
Internal
565 (475)
555 (470)
540 (450)
510 (440)
535 (455)
505 (430)
0.1
NPLC
Digital I/O
490 (420)
470 (405)
470 (410)
450 (390)
470 (400)
450 (390)
1.0
NPLC
Internal
59 (49)
59 (49)
58 (49)
58 (48)
58 (49)
58 (48)
1.0
NPLC
Digital I/O
58 (48)
58 (48)
58 (48)
57 (48)
57 (48)
57 (48)
MAXIMUM MEASUREMENT RANGE CHANGE RATE:
>4500/second typical. When changing to or
from a range
≥
1A, maximum rate is >2000/second typical.
MAXIMUM SOURCE RANGE CHANGE RATE:
>400/second, typical.
MAXIMUM SOURCE FUNCTION CHANGE RATE:
>500/second, typical.
EXTERNAL TRIGGER INPUT:
The Digital I/O interface signals can be configured to behave as trigger
inputs.
Input Latency (time from trigger input to start of measurement or source change):
<150µs, typical.
Input Jitter:
<100µs, typical.
COMMAND PROCESSING TIME:
Maximum time required for the output to begin to change follow-
ing the receipt of the smux.source.levelv or smux.source.leveli command. <1ms typical.
NOTES
1. See the Speed Specifications Test Conditions Appendix in the Series 2600 Reference Manual for more information
regarding test conditions.
Specifications are subject to change without notice. Rev. B