Series 2600 System SourceMeters User’s Manual
Front Panel Operation
1-7
Return to
2600S-900-01 Rev. A / May 2006
1. CHANNEL A and CHANNEL B (Channel B on Models 2602/2612 only)
Input/output connections for source, sense, and guard.
2. DIGITAL I/O
Female DB-25 connector. Fourteen pins for digital input or output, one pin for Output
Enable. Use a cable equipped with a male DB-25 connector (Keithley Instruments part
number CA-126-1CA).
3. IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as the Model
7007-1 or Model 7007-2.
4. Cooling exhaust vent
Exhaust vent for internal cooling fan. Keep vent free of obstructions to prevent
overheating.
5. Chassis ground
Ground screw for connections to chassis ground.
6. Low noise chassis ground
Ground jack for connecting Output HI or LO to chassis.
7. RS-232
Female DB-9 connector. For RS-232 operation, use a straight-through (not null
modem) DB-9 shielded cable for connection to the PC (Keithley Instruments Model 7009-
5).
8. TSP-Link
Expansion interface that allows a Series 2600 and other TSP-enabled instruments to
trigger and communicate with each other. Use a category 5e or higher LAN crossover
cable (Keithley Instruments part number CA-180-3A).
9. Power module
Contains the AC line receptacle and power line fuse. The instrument can operate on
line voltages of 100V to 240VAC at line frequencies of 50Hz or 60Hz. Refer to Section 17
of the Series 2600 Reference Manual for line fuse replacement instructions.