pgu_height
This command sets the peak-to-peak height of the pulse.
Usage
istat = int pgu_height(int
instr_id
, double
height
);
instr_id
The instrument identification code
height
The pulse height in volts (input)
Details
This command returns a
0
if executed without error; a negative number indicates an error.
The pulse height must be greater than -20 V and less than 20 V.
Example
pgu_height(PGU2, 2)
Sets the height of the pulse on PGU2 to 2 V.
Also see
None
pgu_init
This command initializes communication with the pulse card and sets the pulse generator to a specific set of
default conditions.
Usage
istat = int pgu_init(int
instr_id
);
instr_id
The instrument identification code
Details
This command returns a
0
if executed without error; a negative number indicates an error.
The
pgu_init
command sets the pulse generators to the following states:
•
Set to single pulse mode
•
Set output impedance to 50 Ω
•
Set output polarity to normal
•
Enable software triggering
•
Set rise time, fall time to 100 ns
•
Set pulse delay to 0 s
•
Set pulse height to 0.2 V
•
Set pulse width to 500 ns