Section 4: Error definitions
S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual
4-6
S530-900-01 Rev. E / September 2017
650 TAPI_BADCHANNEL
Message
: Request to open unknown channel type XXX.
Remarks
: An internal system software error has occurred.
651 TAPI_BADTESTER
Message
: **This is a host-side only error. No message will be generated by the tester.
Remarks
: There is no network node corresponding to the tester address. Make sure the network
configuration is correct.
652 TAPI_NOTFOUND
Message
: **This is a host-side only error. No message will be generated by the tester.
Remarks
: The tester cannot be located on the network. Make sure the tester is powered on and has
started correctly. This problem can also be caused by network problems.
653 TAPI_REFUSED
Message
: **This is a host-side only error. No message will be generated by the tester.
Remarks
: The tester is in use. This can be caused by another process running diagnostics or a test
plan.
656 TAPI_CHANLIMIT
Message
: Channel limit exceeded.
Remarks
: There are too many active network connections to the tester. Close some of the tools that
communicate with the tester and try again.
657 TAPI_BUFOFLOW
Message
: **This is a host-side only error. No message will be generated by the tester.
Remarks
: An internal system software error has occurred.