Section 3: LPTLib command reference
S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual
3-80
S530-900-01 Rev. E / September 2017
Modifiers that affect CVU measurements
Modifier
Value
Comment
KI_CVU_ACV
<value>
10 mV to 100 mV (45 mV is default)
KI_CVU_ACZ_RANGE
<value>
0
= Autorange (default)
1e-6
= 1 µA range
30e-6
= 30 µA range
1e-3
= 1 mA range
KI_CVU_CORRECT LOAD
0
or
1
0
= Off (default)
1
= On
KI_CVU_CORRECT_OPEN
0
or
1
0
= Off (default)
1
= On
KI_CVU_CORRECT_SHORT
0
or
1
0
= Off (default)
1
= On
KI_CVU_FREQ
<value>
1 kHz to 2 MHz (100 kHz is default)
KI_CVU_LENGTH
KI_CVU_CABLE_CORR_LV
KI_CVU_CABLE_CORR_HV
KI_CVU_CABLE_CORR_LV_HV
Specifies the type of cables used in
the system so that cable-length
corrections can be automated.
KI_CVU_MODE
0
or
1
0
= User mode
1
= System mode (default)
KI_CVU_MODEL
0
to
5
0
= Z, theta
1
= R, jx
2
= Cp, Gp (default)
3
= Cs, Rs
4
= Cp, D
5
= Cs, D
KI_CVU_SPEED
0
to
2
0
= Fast (default)
1
= Normal
2
= Quiet
Does not affect the
intgc
,
intgg
,
or
intgcg
LPT commands. To
adjust the speed of these LPT
commands, use the
KI_INTGPLC
setmode.
Modifiers that affect matrix operation
Modifier
Value
Comment
KI_AUTO_GND_PINS
70xB:
0
or
1
Enable or disable the feature that
automatically connects all unused
pins to GND (including the
CHUCK).
0
= Disable
1
= Enable