Equipment Requirements
6-1
6
Calibration
DAS-1800ST/HR Series boards are initially calibrated at the factory. You
are advised to check the calibration of a board every six months and
calibrate again if necessary. This section provides the information needed
to calibrate a DAS-1800ST/HR Series board.
Equipment Requirements
The equipment requirements for calibrating a DAS-1800ST/HR Series
board are as follows:
●
A digital voltmeter for the DAS-1800ST Series boards accurate to 6
½
digits, such as a Keithley Instruments Model 196.
●
A digital voltmeter for the DAS-1800HR Series boards accurate to
7
½
digits, such as a Keithley Instruments Model 2001.
●
An adjustable ±10 V voltage calibrator, such as a Keithley
Instruments Model 236.
●
An STA-1800U with a CDAS-2000 cable, an STP-50 with a
CDAS-2000 cable, or a user-designed interface.
●
The appropriate number of CDAS-2000 cables for EXP-1800
accessories, if used.