Section 9: Series 3700 Module Schematics and Connections
Series 3700 System Switch/Multimeter User's Manual
9-6
3700S-900-01 Rev. C / July 2008
Slot #
Card
Channel relays closed
Backplane relays closed
SLOT 6
3723
60
4
This produces the following power consumption:
Slot #1 Power Consumed =
700
+
60 x 100
+
4 x 100
=
7100
Slot #2 Power Consumed =
700
+
60 x 100
+
4 x 100
=
7100
Slot #3 Power Consumed =
700
+
60 x 100
+
4 x 100
=
7100
Slot #4 Power Consumed =
700
+
60 x 100
+
4 x 100
=
7100
Slot #5 Power Consumed =
700
+
60 x 100
+
4 x 100
=
7100
Slot #6 Power Consumed =
700
+
60 x 100
+
4 x 100
=
7100
Totals for each bank can then be calculated as follows:
Slot 1
Slot 2
Slot 3
Total
Bank #1 Power Consumed =
7100
+
7100
+ 7100
=
21300
Slot 4
Slot 5
Slot 6
Total
Bank #2 Power Consumed =
7100
+
7100
+ 7100
=
21300
Result: Each bank has exceeded maximum power. Some operations are not
performed and an error is generated.