Section 8: Maintenance
Series 3700 System Switch/Multimeter User's Manual
8-2
3700S-900-01 Rev. C / July 2008
Figure 8-1: Fuse location
Fuse
location
Rating
Keithley Instruments part
number
(1) Analog
backplane
fuse
250V, 3A fast blow 5x20mm
FU-99-1
(2) Instrument
fuse
250V / 1.25A slow blow
5x20mm
FU-106-1.25
To replace a fuse:
1. Using a flat-tip screwdriver, disengage the fuse holder by rotating it counter-
clockwise.
2. Pull out the fuse holder and replace the fuse with the correct type (see
table).
3. Reinstall the fuse holder.
If the fuse continues to blow, a circuit malfunction exists and must be corrected.
Return the unit to Keithley Instruments for repair.
Front panel tests
There are two front panel tests: One to test the functionality of the front panel
keys and one to test the display.